Sistema automático de colector de oxígeno
Fully automatic manifold, the box is made of high-quality 1.0mmsteel plate, and the surface is sprayed.
It contains 2 primarypressure reducing valves, 2 secondary pressure reducing valves,1 globe valve, and 1 24V three-way electric ball valve. Severalvalve joints, 1 three-way ball valve, 3 spare pressure gauges, 10.1touch control screen, 3 pressure sensors.The host has 485 communication, digital display of cylinderpressure, with sound and light alarm, and control the left andright switching through pressure underpressureSupport manual switching of left and right cylinder gas supply,support automatic cylinder gas supply left and right switchingSupport cylinder pressure capacity percentage analog1+1 includes 2 main valves, 1 air outlet valve with adapter, 2 leftand right bus bars plus globe valves, 2 high-pressure hoses, 2brackets, and 2 cylinder protection frames.
Description
1、Left input, right input pressure 0.4~15Mpa
2、Outlet pressure 0.3~1.0Mpa (adjustable)
3、Nominal flow >100M3/h
4、Switching pressure 0.6~1Mpa (can be set)
5、Switching time 3S
6、Working voltage 24VDC
7、Working current 250mA
8、Residual gas pressure of steel cylinder <0.8MPa
9、Kind of Gas: Oxygen
10、High Pressupre Tail's Valve: QF-2; CGA540; (Depending on your demands)